Date on Master's Thesis

12-2025

Document Type

Master's Thesis

Degree Name

M. Eng.

Department

Bioengineering

Committee Chair

Roussel, Thomas

Committee Co-Chair (if applicable)

Voor, Michael

Committee Member

Williams, Stuart J.

Committee Member

Chen, Joseph

Author's Keywords

pedicle screw loosening; cyclic axial loading; stiffness decay; displacement; bone surrogate strength; failure

Abstract

Pedicle screw loosening is a common failure mode in spinal instrumentation, particularly in osteoporotic bone, where cancellous strength averages 0.944 MPa compared to 1.5-7.8 MPa in healthy bone. This study evaluated toggle-out testing for predicting pedicle screw failure and examined how material strength affects failure behavior. Polyurethane blocks underwent progressive cyclic loading from 100 N to 400 N in 100 N increments, with displacement recorded throughout. A 200 N load reflects forces typically carried by a screw during walking in an osteoporotic patient, placing the test range within expected clinical loading. Displacement behavior and stiffness decay were quantified using a power-law model. Weaker blocks (4.9 and 5.85 MPa) exhibited greater displacement accumulation and earlier severe stiffness degradation, while stronger blocks (8.4 and 9.85 MPa) remained stable through higher load levels. These combined metrics identified strength-specific failure thresholds and demonstrated the importance of material strength in predicting screw fixation stability.

Share

COinS